Statements in which the resource exists.
SubjectPredicateObjectContext
pubmed-article:17971345rdf:typepubmed:Citationlld:pubmed
pubmed-article:17971345lifeskim:mentionsumls-concept:C0030705lld:lifeskim
pubmed-article:17971345lifeskim:mentionsumls-concept:C0031809lld:lifeskim
pubmed-article:17971345lifeskim:mentionsumls-concept:C0205476lld:lifeskim
pubmed-article:17971345lifeskim:mentionsumls-concept:C0178602lld:lifeskim
pubmed-article:17971345lifeskim:mentionsumls-concept:C0220825lld:lifeskim
pubmed-article:17971345lifeskim:mentionsumls-concept:C0806487lld:lifeskim
pubmed-article:17971345lifeskim:mentionsumls-concept:C2603343lld:lifeskim
pubmed-article:17971345lifeskim:mentionsumls-concept:C0015737lld:lifeskim
pubmed-article:17971345lifeskim:mentionsumls-concept:C0492899lld:lifeskim
pubmed-article:17971345pubmed:issue4lld:pubmed
pubmed-article:17971345pubmed:dateCreated2008-8-4lld:pubmed
pubmed-article:17971345pubmed:abstractTextThis study presents the results from a survey conducted by the Greek Atomic Energy Commission (GAEC), during the period 1998-2003, in 530 public and private owned fluoroscopic X-ray systems in Greece. Certain operational parameters for conventional and remote control systems were assessed, according to a quality control protocol developed by GAEC on the basis of the current literature. Public (91.5%) and private (81.5%) owned fluoroscopic units exhibit high-contrast resolution values over 1 lp mm(-1). Moreover, 88.5 and 87.1% of the fluoroscopic units installed in the public and private sector, respectively, present Maximum Patient Entrance Kerma Rate values lower than 100 mGy min(-1). Additionally, 68.3% of the units assessed were found to perform within the acceptance limits. Finally, the third quartile of the Entrance Surface Dose Rate distribution was estimated according to the Dose Reference Level definition and found equal to 35 mGy min(-1).lld:pubmed
pubmed-article:17971345pubmed:languageenglld:pubmed
pubmed-article:17971345pubmed:journalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:17971345pubmed:citationSubsetIMlld:pubmed
pubmed-article:17971345pubmed:statusMEDLINElld:pubmed
pubmed-article:17971345pubmed:issn0144-8420lld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:EconomidesSSlld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:HourdakisC...lld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:BoziariAAlld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:KamenopoulouV...lld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:KipourosPPlld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:TritakisPPlld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:KalivasNNlld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:KalathakiMMlld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:SimantirakisG...lld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:ManousaridisG...lld:pubmed
pubmed-article:17971345pubmed:authorpubmed-author:VogiatziSSlld:pubmed
pubmed-article:17971345pubmed:issnTypePrintlld:pubmed
pubmed-article:17971345pubmed:volume129lld:pubmed
pubmed-article:17971345pubmed:ownerNLMlld:pubmed
pubmed-article:17971345pubmed:authorsCompleteYlld:pubmed
pubmed-article:17971345pubmed:pagination419-25lld:pubmed
pubmed-article:17971345pubmed:meshHeadingpubmed-meshheading:17971345...lld:pubmed
pubmed-article:17971345pubmed:meshHeadingpubmed-meshheading:17971345...lld:pubmed
pubmed-article:17971345pubmed:meshHeadingpubmed-meshheading:17971345...lld:pubmed
pubmed-article:17971345pubmed:meshHeadingpubmed-meshheading:17971345...lld:pubmed
pubmed-article:17971345pubmed:meshHeadingpubmed-meshheading:17971345...lld:pubmed
pubmed-article:17971345pubmed:meshHeadingpubmed-meshheading:17971345...lld:pubmed
pubmed-article:17971345pubmed:meshHeadingpubmed-meshheading:17971345...lld:pubmed
pubmed-article:17971345pubmed:meshHeadingpubmed-meshheading:17971345...lld:pubmed
pubmed-article:17971345pubmed:year2008lld:pubmed
pubmed-article:17971345pubmed:articleTitleImage quality evaluation and patient dose assessment of medical fluoroscopic X-ray systems: a national study.lld:pubmed
pubmed-article:17971345pubmed:affiliationGreek Atomic Energy Commission, Ag. Paraskevi, Greece.lld:pubmed
pubmed-article:17971345pubmed:publicationTypeJournal Articlelld:pubmed
pubmed-article:17971345pubmed:publicationTypeEvaluation Studieslld:pubmed