This study presents the results from a survey conducted by the Greek Atomic Energy Commission (GAEC), during the period 1998-2003, in 530 public and private owned fluoroscopic X-ray systems in Greece. Certain operational parameters for conventional and remote control systems were assessed, according to a quality control protocol developed by GAEC on the basis of the current literature. Public (91.5%) and private (81.5%) owned fluoroscopic units exhibit high-contrast resolution values over 1 lp mm(-1). Moreover, 88.5 and 87.1% of the fluoroscopic units installed in the public and private sector, respectively, present Maximum Patient Entrance Kerma Rate values lower than 100 mGy min(-1). Additionally, 68.3% of the units assessed were found to perform within the acceptance limits. Finally, the third quartile of the Entrance Surface Dose Rate distribution was estimated according to the Dose Reference Level definition and found equal to 35 mGy min(-1).
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:issn |
0144-8420
|
pubmed:author |
pubmed-author:BoziariAA,
pubmed-author:EconomidesSS,
pubmed-author:HourdakisC JCJ,
pubmed-author:KalathakiMM,
pubmed-author:KalivasNN,
pubmed-author:KamenopoulouVV,
pubmed-author:KipourosPP,
pubmed-author:ManousaridisGG,
pubmed-author:SimantirakisGG,
pubmed-author:TritakisPP,
pubmed-author:VogiatziSS
|
pubmed:issnType |
Print
|
pubmed:volume |
129
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
419-25
|
pubmed:meshHeading |
pubmed-meshheading:17971345-Fluoroscopy,
pubmed-meshheading:17971345-Greece,
pubmed-meshheading:17971345-Humans,
pubmed-meshheading:17971345-Private Sector,
pubmed-meshheading:17971345-Public Sector,
pubmed-meshheading:17971345-Quality Control,
pubmed-meshheading:17971345-Radiation Dosage,
pubmed-meshheading:17971345-Radiographic Image Interpretation, Computer-Assisted
|
pubmed:year |
2008
|
pubmed:articleTitle |
Image quality evaluation and patient dose assessment of medical fluoroscopic X-ray systems: a national study.
|
pubmed:affiliation |
Greek Atomic Energy Commission, Ag. Paraskevi, Greece.
|
pubmed:publicationType |
Journal Article,
Evaluation Studies
|