This study presents the results from a survey conducted by the Greek Atomic Energy Commission (GAEC), during the period 1998-2003, in 530 public and private owned fluoroscopic X-ray systems in Greece. Certain operational parameters for conventional and remote control systems were assessed, according to a quality control protocol developed by GAEC on the basis of the current literature. Public (91.5%) and private (81.5%) owned fluoroscopic units exhibit high-contrast resolution values over 1 lp mm(-1). Moreover, 88.5 and 87.1% of the fluoroscopic units installed in the public and private sector, respectively, present Maximum Patient Entrance Kerma Rate values lower than 100 mGy min(-1). Additionally, 68.3% of the units assessed were found to perform within the acceptance limits. Finally, the third quartile of the Entrance Surface Dose Rate distribution was estimated according to the Dose Reference Level definition and found equal to 35 mGy min(-1).
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0144-8420
pubmed:author
pubmed:issnType
Print
pubmed:volume
129
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
419-25
pubmed:meshHeading
pubmed:year
2008
pubmed:articleTitle
Image quality evaluation and patient dose assessment of medical fluoroscopic X-ray systems: a national study.
pubmed:affiliation
Greek Atomic Energy Commission, Ag. Paraskevi, Greece.
pubmed:publicationType
Journal Article, Evaluation Studies