pubmed:abstractText
Surface plasmon resonance imaging systems, such as Flexchip from Biacore, are capable of monitoring hundreds of reaction spots simultaneously within a single flow cell. Interpreting the binding kinetics in a large-format flow cell presents a number of potential challenges, including accounting for mass transport effects and spot-to-spot sample depletion. We employed a combination of computer simulations and experimentation to characterize these effects across the spotted array and established that a simple two-compartment model may be used to accurately extract intrinsic rate constants from the array under mass transport-limited conditions. Using antibody systems, we demonstrate that the spot-to-spot variability in the binding kinetics was <9%. We also illustrate the advantage of globally fitting binding data from multiple spots within an array for a system that is mass transport limited.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Feb
pubmed:issn
0003-2697
pubmed:author
pubmed:issnType
Print
pubmed:day
1
pubmed:volume
373
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
112-20
pubmed:meshHeading
pubmed:year
2008
pubmed:articleTitle
Extracting kinetic rate constants from surface plasmon resonance array systems.
pubmed:affiliation
Center for Biomolecular Interaction Analysis, School of Medicine, University of Utah, Salt Lake City, UT 84132, USA.
pubmed:publicationType
Journal Article, Research Support, U.S. Gov't, Non-P.H.S.