pubmed:abstractText
The main cause of cardiac valve bioprosthesis dysfunction is mechanical rupture of the cusps consequent upon the development of "fatigue" tensions and calcification of the biologic tissue. The technology of "Bionix" prostheses manufacturing implies prevention of these complications by means of supporting frameworks of variable rigidity (dumping stroke loads) and the use of a calcinosis inhibitor in the process of chemical stabilization. Experimental research confirms the certain effectiveness of such an approach to solving the problem of bioprosthesis durability. Experience in clinical application of biological valves of this series showed that by the 5th postoperative year the stability of good results was 90.0 +/- 2.2%; absence of thromboembolic complications in 93.3 +/- 2.4%, and no need to operate because of dysfunction of the bioprosthesis in 96.6 +/- 2.1%.
pubmed:language
rus
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
0236-2791
pubmed:author
pubmed:issnType
Print
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
11-6
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1991
pubmed:articleTitle
[Heart valve bioprostheses of the "Bionix" series. 6-year experience with the mitral valve replacement].