pubmed:abstractText |
This acute feasibility study compared two different automatic capture detection methodologies, the reduced coupling capacitor (RCC) and the independent pace/sense (IPS) methods, for the left ventricle (LV).
|
pubmed:language |
eng
|
pubmed:journal |
|
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jul
|
pubmed:issn |
0147-8389
|
pubmed:author |
|
pubmed:issnType |
Print
|
pubmed:volume |
30
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
865-73
|
pubmed:meshHeading |
pubmed-meshheading:17584268-Cardiac Pacing, Artificial,
pubmed-meshheading:17584268-Evoked Potentials,
pubmed-meshheading:17584268-Feasibility Studies,
pubmed-meshheading:17584268-Female,
pubmed-meshheading:17584268-Humans,
pubmed-meshheading:17584268-Male,
pubmed-meshheading:17584268-Middle Aged,
pubmed-meshheading:17584268-Pacemaker, Artificial,
pubmed-meshheading:17584268-Prospective Studies,
pubmed-meshheading:17584268-Ventricular Dysfunction, Left
|
pubmed:year |
2007
|
pubmed:articleTitle |
Clinical evaluation of two different evoked response sensing methods for automatic capture detection in the left ventricle.
|
pubmed:affiliation |
Deutsches Herzzentrum Berlin, Berlin, Germany. goetze@dhzb.de
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Research Support, Non-U.S. Gov't,
Evaluation Studies
|