2007-2-27
|
pubmed:abstractText |
Discrepancies between serum and heparin plasma samples have been described for many commercial troponin assays including the cardiac troponin T (cTnT) assay. Using the current 3rd generation Elecsys Troponin T immunoassay, heparin plasma cannot be recommended for the determination of cTnT due to systematic lower test results caused by a direct interference of the immunoassay by heparin. The purpose of the multicenter study was to evaluate the analytical performance of an improved 4th generation Elecsys Troponin T immunoassay with a special focus on the comparability of cTnT results determined in heparin plasma and serum.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:issn |
1433-6510
|
pubmed:author |
pubmed-author:AppleFF,
pubmed-author:Garcia-BeltrànLL,
pubmed-author:HermsenDD,
pubmed-author:JaffeAA,
pubmed-author:JarauschJJ,
pubmed-author:KaronBB,
pubmed-author:LewandrowskiEE,
pubmed-author:MühlbacherAA,
pubmed-author:MüllerRR,
pubmed-author:OrdóñezJJ,
pubmed-author:PaganiFF,
pubmed-author:PanteghiniMM,
pubmed-author:PleckoTT
|
pubmed:issnType |
Print
|
pubmed:volume |
53
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1-9
|
pubmed:dateRevised |
2007-11-15
|
pubmed:meshHeading | |
pubmed:year |
2007
|
pubmed:articleTitle |
Results from a multicenter evaluation of the 4th generation Elecsys Troponin T assay.
|
pubmed:affiliation |
Institute of Clinical Chemistry and Laboratorv Diagnostics, Heinrich-Heine University, Düsseldorf, Germany. hermsen@med.uni-duesseldorf.de
|