pubmed:abstractText
Inappropriate shock remains a major issue in patients with implantable cardioverter defibrillators. We examine an inappropriate shock delivered in a patient with atrial tachycardia that conducted 1:1. We reconstruct the device algorithms that led to therapy delivery and discuss programming changes that could be considered.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
0147-8389
pubmed:author
pubmed:issnType
Print
pubmed:volume
29
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1413-5
pubmed:meshHeading
pubmed:year
2006
pubmed:articleTitle
Inappropriate shock: a failure of SVT discriminators in a dual chamber ICD?
pubmed:affiliation
Cardiac Arrhythmia Service, Stanford University School of Medicine, Stanford, California 94305, USA. aalahmad@cvmed.stanford.edu