Patients with implantable defibrillators often require bradycardia pacemakers. Adverse interactions between separate defibrillator and bradycardia pacing units have occurred, including failure to detect ventricular fibrillation due to persistent bradycardia pacing during the arrhythmia. A device with combined bradycardia pacing and antitachycardia therapy capability may obviate adverse device interactions. We describe a previously unrecognized phenomenon that may occur in a combined device when the algorithms for sensing bradycardia and tachycardia are "codependent"; that is, the circuitry for brady- and tachyarrhythmia detection relies on the same automatic gain sense amplifier. Three of 37 patients in whom the device was implanted had ventricular tachycardia initiated when bradycardia pacing stimuli were delivered by the device after probable nonsensed sinus beats. In each case, nonsensed beats appeared to have a markedly diminished amplitude, occurred after ventricular premature depolarizations that produced large amplitude electrograms, and had an electrogram morphology that matched that of sinus rhythm. In each case, the bradycardia pacing interval was at least 1,200 msec (range 1,200 to 1,714 msec). In two of the three patients, large amplitude ventricular premature depolarizations or nonsustained ventricular tachycardia caused an adjustment of the gain control that potentiated the failure to sense the subsequent lower amplitude signal. In all three patients, the induced arrhythmia was rapidly terminated by pacing or cardioversion. Decreasing the bradycardia pacing interval by 110-514 msec has prevented recurrence during short-term follow-up. Our findings suggest that codependent bradycardia and antitachycardia devices may have their own unique potential difficulties in adapting to rapid changes in rate and signal amplitude.
pubmed:grant
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Aug
pubmed:issn
0147-8389
pubmed:author
pubmed:issnType
Print
pubmed:volume
14
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1281-7
pubmed:dateRevised
2007-11-14
pubmed:meshHeading
pubmed:year
1991
pubmed:articleTitle
Paced beats following single nonsensed complexes in a "codependent" cardioverter defibrillator and bradycardia pacing system: potential for ventricular tachycardia induction.
pubmed:affiliation
Clinical Electrophysiology Laboratory, Hospital of the University of Pennsylvania, Philadelphia 19104.
pubmed:publicationType
Journal Article, Research Support, U.S. Gov't, P.H.S., Case Reports, Research Support, Non-U.S. Gov't