Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
4
pubmed:dateCreated
2006-12-15
pubmed:abstractText
The development of a multielectrode array is the key issue for retinal prostheses. We developed a 10 x 10 platinum electrode array that consists of an 8-microm polyimide layer sandwiched between 5-microm polymonochloro-para-xylylene (parylene-C) layers. Each electrode was formed as a 30-microm-high bump by Pt/Au double-layer electroplating. We estimated the charge delivery capability (CDC) of the electrode by measuring the CDCs of two-channel electrode arrays. The dimensions of each electrode of the two-channel array were the same as those of each electrode formed on the 10 x 10 array. The results suggest that for cathodic-first (CF) pulses, 80% of electrodes surpassed our development target of 318 microC/cm2, which corresponds to the charge density of pulses of 500 micros duration and 200 microA amplitude for a 200-microm-diameter planar electrode.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
1434-7229
pubmed:author
pubmed:issnType
Print
pubmed:volume
9
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
263-6
pubmed:meshHeading
pubmed:year
2006
pubmed:articleTitle
The development of a multichannel electrode array for retinal prostheses.
pubmed:affiliation
Vision Institute, Nidek Co, Ltd, Gamagori 443-0036, Japan. yasuo_terasawa@nidek.co.jp
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't