Source:http://linkedlifedata.com/resource/pubmed/id/16949894
Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
1-2
|
pubmed:dateCreated |
2007-1-29
|
pubmed:abstractText |
The first HPLC-fluorescence method for the determination of atomoxetine in human plasma was developed and validated. Atomoxetine was derivatized with 4-(4,5-diphenyl-1H-imidazol-2-yl) benzoyl chloride (DIB-Cl) under mild conditions, and separated isocratically on a C18 column using a HPLC system with fluorescence detection (lambdaex: 318 nm, lambdaem: 448 nm). A linear calibration curve was obtained over the concentration range 1-1000 ng/mL (r=0.999). The limit of detection (S/N=3) was 0.3 ng/mL. The relative standard deviations of intra-day and inter-day variations were < or =8.30% and 7.47%, respectively. This method is rapid, sensitive, and suitable for both basic and clinical studies of atomoxetine.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical |
http://linkedlifedata.com/resource/pubmed/chemical/4-(4,5-diphenyl-1H-imidazol-2-yl)ben...,
http://linkedlifedata.com/resource/pubmed/chemical/Adrenergic Uptake Inhibitors,
http://linkedlifedata.com/resource/pubmed/chemical/Benzoates,
http://linkedlifedata.com/resource/pubmed/chemical/Imidazoles,
http://linkedlifedata.com/resource/pubmed/chemical/Propylamines,
http://linkedlifedata.com/resource/pubmed/chemical/atomoxetine
|
pubmed:status |
MEDLINE
|
pubmed:month |
Feb
|
pubmed:issn |
1570-0232
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:day |
1
|
pubmed:volume |
846
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
351-4
|
pubmed:dateRevised |
2007-11-15
|
pubmed:meshHeading |
pubmed-meshheading:16949894-Adrenergic Uptake Inhibitors,
pubmed-meshheading:16949894-Benzoates,
pubmed-meshheading:16949894-Chromatography, High Pressure Liquid,
pubmed-meshheading:16949894-Humans,
pubmed-meshheading:16949894-Imidazoles,
pubmed-meshheading:16949894-Propylamines,
pubmed-meshheading:16949894-Reproducibility of Results,
pubmed-meshheading:16949894-Sensitivity and Specificity,
pubmed-meshheading:16949894-Spectrometry, Fluorescence
|
pubmed:year |
2007
|
pubmed:articleTitle |
Sensitive quantification of atomoxetine in human plasma by HPLC with fluorescence detection using 4-(4,5-diphenyl-1H-imidazole-2-yl) benzoyl chloride derivatization.
|
pubmed:affiliation |
Laboratory of Drug Disposition and Pharmacogenetics, Department of Pharmaceutical Sciences, Medical University of South Carolina, Charleston, SC 29425, USA.
|
pubmed:publicationType |
Journal Article,
Validation Studies
|