The purpose of this study was to determine if electronic apex locators (EAL) or electric pulp testers (EPT) interfere with the function of implanted cardiac pacemakers (ICP) or cardioverter/defibrillators (ICD). Twenty-seven patients with ICPs or ICDs had continuous electrocardiogram monitoring and device interrogation to detect interferences during the use of two types of EALs and one EPT. No interferences were detected by any ICP or ICD. In six patients, with intermittent pacing, a significant increase in pacing was observed during EAL/EPT stimulation (p < 0.05). Examination of RR intervals (a measure of intrinsic heart rate) demonstrated significantly longer RR intervals (slower intrinsic heart rate) during EAL/EPT stimulation (p < 0.05). Evaluation of the electrocardiograms for each patient failed to show any abnormalities in pacing during testing. These findings led us to conclude that the increased pacing frequency observed was related to a slower intrinsic heart rate and not electrical interference with the cardiac devices. In conclusion, the two EALs and one EPT used in this study did not interfere with the functioning of any of the cardiac devices tested.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
D
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0099-2399
pubmed:author
pubmed:issnType
Print
pubmed:volume
32
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
847-52
pubmed:meshHeading
pubmed:year
2006
pubmed:articleTitle
Safety of electronic apex locators and pulp testers in patients with implanted cardiac pacemakers or cardioverter/defibrillators.
pubmed:affiliation
Department of Endodontology, School of Dentistry, Oregon Health & Science University, Portland, OR 97201, USA. dr.brianwilson@comcast.net
pubmed:publicationType
Journal Article