Underlying characteristics in brachytherapy dosimetry parameters for medical radionuclides 137Cs, 125I, 192Ir, 103Pd, and 169Yb were examined using Monte Carlo methods. Sources were modeled as unencapsulated point or line sources in liquid water to negate variations due to materials and construction. Importance of phantom size, mode of radiation transport physics--i.e., photon transport only or coupled photon:electron transport, phantom material, volume averaging, and Monte Carlo tally type were studied. For noninfinite media, g(r) was found to degrade as r approached R, the phantom radius. MCNP5 results were in agreement with those published using GEANT4. Brachytherapy dosimetry parameters calculated using coupled photon:electron radiation transport simulations did not differ significantly from those using photon transport only. Dose distributions from low-energy photon-emitting radionuclides 125I and 103Pd were sensitive to phantom material by upto a factor of 1.4 and 2.0, respectively, between tissue-equivalent materials and water at r =9 cm. In comparison, high-energy photons from 137Cs, 192Ir, and 169Yb demonstrated +/- 5% differences in dose distributions between water and tissue substitutes at r=20 cm. Similarly, volume-averaging effects were found to be more significant for low-energy radionuclides. When modeling line sources with L < or = 0.5 cm, the two-dimensional anisotropy function was largely within +/- 0.5% of unity for 137Cs, 125I, and 192Ir. However, an energy and geometry effect was noted for 103Pd and 169Yb, with Pd-103F(0.5,0 degrees)=l.05 and yb-169F(0.5,0 degrees)=0.98 for L=0.5 cm. Simulations of monoenergetic photons for L=0.5 cm produced energy-dependent variations in F(r, theta) having a maximum value at 10 keV, minimum at 50 keV, and approximately 1.0 for higher-energy photons up to 750 keV. Both the F6 cell heating and *F4 track-length estimators were employed to determine brachytherapy dosimetry parameters. F6 was found to be necessary for g(r), while both tallies provided equivalent results for F(r, theta).
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0094-2405
pubmed:author
pubmed:issnType
Print
pubmed:volume
33
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1729-37
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2006
pubmed:articleTitle
Approaches to calculating AAPM TG-43 brachytherapy dosimetry parameters for 137Cs, 125I, 192Ir, 103Pd, and 169Yb sources.
pubmed:affiliation
Department of Radiation Oncology, Tufts University School of Medicine, Boston, Massachusetts 02111, USA.
pubmed:publicationType
Journal Article, Comparative Study