2006-5-11
|
pubmed:abstractText |
The first clinical examination using a two-dimensional imaging system of coronary angiography with monochromated synchrotron radiation was carried out at the National Laboratory for High Energy Physics in May 1996. A rotating X-ray shutter was developed to produce a pulsed X-ray beam with 2-6 ms of beam spill to suppress image blurring, at a frequency of 30 Hz. A performance test of the X-ray shutter using synchrotron radiation was carried out, and it was verified that the shutter had satisfactory specifications for clinical applications. With this X-ray shutter the monitored radiation dose in clinical examinations was consistent with theoretical expectations and kept within a reasonable level of radiation protection.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:status |
PubMed-not-MEDLINE
|
pubmed:month |
Mar
|
pubmed:issn |
0909-0495
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:day |
1
|
pubmed:volume |
5
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
123-7
|
pubmed:year |
1998
|
pubmed:articleTitle |
Development of a rotating X-ray shutter for coronary angiography using synchrotron radiation.
|