pubmed:abstractText
Microbial contamination of dental chair unit (DCU) output water caused by biofilm growth in dental unit waterlines (DUWs) is a universal problem and a potentially significant source of cross-infection. The microbial quality of output water from a Planmeca Compact i DCU equipped with the novel Water Management System (WMS), an integrated and automated DUW cleaning system, was investigated over a 12-month period with the hydrogen peroxide- and silver ion-containing disinfectants Planosil and Planosil Forte. Four weeks after connection to the potable-water quality mains supply the density of aerobic heterotrophic bacteria, rose from the low levels consistently found in the supply water throughout this study (mean average 77 cfu/mL) to 15,400 cfu/mL. Disinfection of DUWs once weekly with Planosil for 10 weeks resulted in a dramatic reduction in bacterial density immediately following disinfection (mean average 26 cfu/mL). Bacterial density rose steadily between disinfections and by 7 days post-disinfection, water quality failed (mean average 384 cfu/mL) the American Dental Association DCU water quality standard of <or=200 cfu/mL. The DCU was then disinfected once weekly for 40 weeks with Planosil Forte. The average bacterial density immediately post-disinfection was 20 cfu/mL and 7 days post-disinfection was 113 cfu/mL. Electron microscopy showed that improved output water quality following disinfection with both disinfectants was associated with marked elimination of DUW biofilm, but deterioration of water quality following disinfection was associated with its regrowth. The most common bacterial species cultured from the mains water and the DCU output water were Microcococcus luteus and Sphingomonas spp., respectively, the latter of which are known opportunistic pathogens. The findings of this study show that the Planmeca Compact i DCU equipped with the easy to use and automated WMS, that requires minimal effort on the part of the operator, consistently provides output water that passes the ADA quality standard of <or=200 cfu/mL for up to 7 days following once-weekly disinfection with Planosil Forte.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
D
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
0300-5712
pubmed:author
pubmed:issnType
Print
pubmed:volume
34
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
648-61
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2006
pubmed:articleTitle
A novel automated waterline cleaning system that facilitates effective and consistent control of microbial biofilm contamination of dental chair unit waterlines: a one-year study.
pubmed:affiliation
Microbiology Research Unit, Division of Oral Biosciences, Dublin Dental School & Hospital, University of Dublin, Trinity College, Lincoln Place, Dublin 2, Republic of Ireland.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't, Evaluation Studies