To evaluate the applicability of dual energy x-ray bone absorptiometry in newborns, precision and accuracy of the method was tested for very small quantities of mineral ranging from less than 0.5 g to about 4 g of hydroxyapatite using a Hologic QDR 1000 instrument. For six femurs excised from preterm stillborns, the mean precision for bone mineral content (BMC) and bone mineral density (BMD) was 1.2 and 0.8%, respectively. Accuracy based on ash weight indicated a mean overestimation of about 7%. In vivo, the precision was assessed by measuring lumbar spine BMC and BMD (L1 to L5) two or three times in 10 newborns (gestational age, 33 to 40 wk). The mean coefficients of variation were 2.40 and 1.55% for BMC and BMD, respectively. Lumbar spine BMC and BMD were also measured once in 30 full-term infants. Values ranged from 1.17 to 3.90 g for BMC and from 0.192 to 0.356 g/cm2 for BMD. The present study shows that dual energy x-ray absorptiometry provides a valuable new tool for the assessment and management of BMC in low birth weight infants and neonates in general.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jul
pubmed:issn
0031-3998
pubmed:author
pubmed:issnType
Print
pubmed:volume
32
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
77-80
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1992
pubmed:articleTitle
Dual energy x-ray absorptiometry measurement of bone mineral content in newborns: validation of the technique.
pubmed:affiliation
INSERM Unit 234, Edouard Herriot Hospital, Lyon, France.
pubmed:publicationType
Journal Article