pubmed:abstractText |
RAISUS is a system for rapid bacterial identification and antimicrobial susceptibility testing. RAISUS and VITEK showed 97.8% and 75.9% agreement in identification of 45 Staphylococcus aureus strains and 58 coagulase-negative staphylococci (CoNS), respectively, and RAISUS and CLSI (formerly NCCLS) methods showed 87.2% and 87.9% agreement in the MICs for S. aureus and CoNS, respectively. RAISUS provided these data within 3.75 h, suggesting its utility for clinical bacteriological laboratories.
|
pubmed:commentsCorrections |
|
pubmed:language |
eng
|
pubmed:journal |
|
pubmed:citationSubset |
IM
|
pubmed:chemical |
|
pubmed:status |
MEDLINE
|
pubmed:month |
Nov
|
pubmed:issn |
0095-1137
|
pubmed:author |
|
pubmed:issnType |
Print
|
pubmed:volume |
43
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
5808-10
|
pubmed:dateRevised |
2009-11-18
|
pubmed:meshHeading |
pubmed-meshheading:16272527-Anti-Bacterial Agents,
pubmed-meshheading:16272527-Automation,
pubmed-meshheading:16272527-Enzymes,
pubmed-meshheading:16272527-Evaluation Studies as Topic,
pubmed-meshheading:16272527-Fluorescence,
pubmed-meshheading:16272527-Humans,
pubmed-meshheading:16272527-Microbial Sensitivity Tests,
pubmed-meshheading:16272527-Oxidation-Reduction,
pubmed-meshheading:16272527-Sensitivity and Specificity,
pubmed-meshheading:16272527-Staphylococcal Infections,
pubmed-meshheading:16272527-Staphylococcus aureus
|
pubmed:year |
2005
|
pubmed:articleTitle |
Evaluation of a fully automated system (RAISUS) for rapid identification and antimicrobial susceptibility testing of Staphylococci.
|
pubmed:affiliation |
Department of Infection Control and Laboratory Diagnostics, Tohoku University Graduate School of Medicine, 1-1 Seiryo-machi, Aobaku, Sendai, Miyagi 980-8574, Japan. kane2@mail.tains.tohoku.ac.jp
|
pubmed:publicationType |
Journal Article,
Comparative Study
|