Detector systems using plastic scintillators can provide instantaneous measurements with high spatial resolution in many applications including small field and high dose gradient field applications. Energy independence and water equivalence are important dosimetric properties that determine whether a detector will be useful in a clinical setting. Using Monte Carlo simulations, we calculated the energy dependence of plastic scintillators when exposed to photon beams in the radiotherapeutic range. These calculations were performed for a detector comprised of a BC-400 plastic scintillator surrounded by a polystyrene wall. Our results showed the plastic scintillation detector to be nearly energy independent over a range of energies from 0.5 to 20 MeV. The ratio of the dose absorbed by the scintillator to that absorbed by water was nearly a constant, approximately equal to 0.98 over the entire energy range of interest. These results confirm the water equivalence of the plastic scintillation detector and are in very good agreement with earlier results obtained using Burlin cavity theory.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
May
pubmed:issn
0094-2405
pubmed:author
pubmed:issnType
Print
pubmed:volume
32
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1265-9
pubmed:meshHeading
pubmed:year
2005
pubmed:articleTitle
Monte Carlo calculations of the absorbed dose and energy dependence of plastic scintillators.
pubmed:affiliation
Department of Radiation Physics, Division of Radiation Oncology, The University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030, USA. abeddar@mdanderson.org
pubmed:publicationType
Journal Article