pubmed:abstractText
In 2001, the Committee on Standardization of Laboratory Testing Related to Diabetes Mellitus of the Japan Diabetes Society (JDS) prepared and certified a new reference material for haemoglobin A1c (HbA1c), Lot 2. The standardization programme for HbA1c measurement in Japan is currently based on Lot 2, although some laboratories still use the previous material (Lot 1). The values assigned to Lot 2 were based on the consensus values for Lot 1 and should give the same results. Therefore, there should be no difference in the measured values no matter which calibrators are used. The Committee conducted a domestic survey in order to confirm this relationship.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
0004-5632
pubmed:author
pubmed:issnType
Print
pubmed:volume
42
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
47-50
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2005
pubmed:articleTitle
Japanese standard reference material JDS Lot 2 for haemoglobin A1c. II: Present state of standardization of haemoglobin A1c in Japan using the new reference material in routine clinical assays.
pubmed:affiliation
Department of Laboratory Medicine, Yamagata University School of Medicine, 2-2-2 Iida-Nishi, Yamagata 990-9585, Japan. mtominag@med.id.yamagata-u.ac.jp
pubmed:publicationType
Journal Article, Comparative Study