In spoiled gradient echo sequences, the T(1)-weighting of image contrast is strongly affected by a nonlinear interaction of two sequence parameters, repetition time (TR) and flip angle (alpha). If alpha is not properly adjusted to compensate for variation in TR due to changing resolution, bandwidth, or number of slices, any optimization of contrast-to-noise may be compromised. Currently, there is no direct way to compare or reproduce the contrast properties of one sequence to another with a different TR. Here, it is demonstrated that for short TRs alpha may be calculated and automatically adjusted such that relative contrast--the shape of the signal-versus-T(1) curve--remains consistent and signal scales proportionally to radicalTR in all tissues. TR is then free to vary to accommodate a range of sequence parameters without impacting relative contrast and the T(1)-weighting of one sequence can be compared to, or reproduced in, another study with a different TR.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0740-3194
pubmed:author
pubmed:copyrightInfo
Copyright 2005 Wiley-Liss, Inc.
pubmed:issnType
Print
pubmed:volume
53
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
977-80
pubmed:meshHeading
pubmed:year
2005
pubmed:articleTitle
Flip angle calculation for consistent contrast in spoiled gradient echo imaging.
pubmed:affiliation
GE Healthcare, Applied Science Lab, 333 Ravenswood Ave, Building 307, Menlo Park, California, USA. reed.busse@med.ge.com
pubmed:publicationType
Journal Article