pubmed:abstractText |
Intraaortic balloon counterpulsation (IABP) timing errors during arrhythmia may result in afterload increases which may negatively influence left ventricular (LV) ejection and LV mechanical dyssynchrony. The aim of our study was to determine beat-to-beat effects of properly timed IABP, premature IAB inflation, and late IAB deflation on LV performance and LV mechanical dyssynchrony in heart failure patients undergoing cardiac surgery.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
AIM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Mar
|
pubmed:issn |
1552-6259
|
pubmed:author | |
pubmed:issnType |
Electronic
|
pubmed:volume |
79
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
872-80
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading | |
pubmed:year |
2005
|
pubmed:articleTitle |
Beat-to-beat effects of intraaortic balloon pump timing on left ventricular performance in patients with low ejection fraction.
|