Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1
pubmed:dateCreated
2005-2-4
pubmed:abstractText
Transepidermal water loss (TEWL) is one of the most important biophysical parameters for evaluating the efficiency of the human skin water barrier. Different approaches exist to measure TEWL. The most commonly used methodology consists of the open chamber diffusion technique in which the water vapor pressure gradient is measured in g/h m2 according to Fick's law. A typical apparatus is the Tewameter. Recently, a portable device--the VapoMeter--became available with a humidity sensor in a closed chamber.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Feb
pubmed:issn
0909-752X
pubmed:author
pubmed:issnType
Print
pubmed:volume
11
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
61-9
pubmed:dateRevised
2008-11-21
pubmed:meshHeading
pubmed:year
2005
pubmed:articleTitle
Validation of the VapoMeter, a closed unventilated chamber system to assess transepidermal water loss vs. the open chamber Tewameter.
pubmed:affiliation
Department of Toxicology, Dermato-Cosmetology and Pharmacognosy, Vrije Universiteit Brussel, Brussels, Belgium. kdepaepe@fafy.vuc.ac.be
pubmed:publicationType