pubmed:abstractText |
Opinions vary regarding the need to perform defibrillation testing prior to hospital discharge in recipients of state-of-the-art cardioverter defibrillators (ICDs). Our protocol is to perform predischarge ICD testing 1 day after implant. This report includes 682 consecutive implants. Adverse observations at testing were grouped into (1) risk of defibrillation failure, (2) surgical complications, (3) sensing/pacing issues or narrow defibrillation margin warranting closer follow-up, or (4) findings correctable by device reprogramming. Among the 682 patients, 63% had single-chamber and 37% dual-chamber or biventricular ICDs. In 48 patients (7%) there were 69 concerns and/or interventions, with overlaps among the four categories, including one failure to defibrillate (0.15%), and six other patients at risk. Surgical complications included 11 hematomas (1.6%), and six lead dysfunctions. Closer follow-up was indicated in 19 patients (2.7%), for high pacing thresholds in seven, sensing issues in seven, and <10 J defibrillation margin in five. Device reprogramming was needed in 31 patients (4.5%), for tachycardia detection and therapy settings in 12, and for pacing/sensing functions in 22 patients. In eight patients ventricular fibrillation could not be induced. There was no morbidity or mortality due to testing. The state-of-the-art ICDs delivering biphasic shocks are remarkably reliable. The routine pre-hospital discharge defibrillation testing of such ICDs may be optional and left to the physicians' discretion.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jan
|
pubmed:issn |
0147-8389
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
28 Suppl 1
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
S260-2
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading | |
pubmed:year |
2005
|
pubmed:articleTitle |
Value of pre-hospital discharge defibrillation testing in recipients of implanted cardioverter defibrillators.
|
pubmed:affiliation |
Division of Cardiology, Department of Medicine, Arrhythmia Service, Montefiore Medical Center, Bronx, NY 10467, USA.
|
pubmed:publicationType |
Journal Article,
Research Support, Non-U.S. Gov't
|