pubmed:abstractText
Automatic Sensitivity Control, unique to St. Jude defibrillators, allows for programmable sensitivity threshold and refractory periods. However, inappropriate programming can lead to marked undersensing and potentially lethal complications. We report a case of complete failure to sense ventricular fibrillation with inappropriate sensitivity adjustments. Detailed review of the parameters with induction of ventricular fibrillation must be performed to assure adequate safety.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0147-8389
pubmed:author
pubmed:issnType
Print
pubmed:volume
27
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1321-3
pubmed:meshHeading
pubmed:year
2004
pubmed:articleTitle
Total failure to sense ventricular fibrillation with inappropriate defibrillator sensitivity adjustment.
pubmed:affiliation
Electrophysiology Service, Cardiovascular Division, University of Pennsylvania Health System, Philadelphia, Pennsylvania 19104, USA.
pubmed:publicationType
Journal Article, Case Reports