pubmed:abstractText |
Automatic Sensitivity Control, unique to St. Jude defibrillators, allows for programmable sensitivity threshold and refractory periods. However, inappropriate programming can lead to marked undersensing and potentially lethal complications. We report a case of complete failure to sense ventricular fibrillation with inappropriate sensitivity adjustments. Detailed review of the parameters with induction of ventricular fibrillation must be performed to assure adequate safety.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Sep
|
pubmed:issn |
0147-8389
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
27
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1321-3
|
pubmed:meshHeading |
pubmed-meshheading:15461727-Aged,
pubmed-meshheading:15461727-Defibrillators, Implantable,
pubmed-meshheading:15461727-Equipment Design,
pubmed-meshheading:15461727-Equipment Failure,
pubmed-meshheading:15461727-Female,
pubmed-meshheading:15461727-Humans,
pubmed-meshheading:15461727-Ventricular Fibrillation
|
pubmed:year |
2004
|
pubmed:articleTitle |
Total failure to sense ventricular fibrillation with inappropriate defibrillator sensitivity adjustment.
|
pubmed:affiliation |
Electrophysiology Service, Cardiovascular Division, University of Pennsylvania Health System, Philadelphia, Pennsylvania 19104, USA.
|
pubmed:publicationType |
Journal Article,
Case Reports
|