Source:http://linkedlifedata.com/resource/pubmed/id/15456316
Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
19
|
pubmed:dateCreated |
2004-9-30
|
pubmed:abstractText |
A novel method for the detection of arsenic(III) in 1 M HCl at a gold nanoparticle-modified glassy carbon electrode has been developed. The gold nanoparticles were electrodeposited onto the glassy carbon electrode via a potential step from +1.055 to -0.045 V vs SCE for 15 s from 0.5 M H2SO4 containing 0.1 mM HAuCl4. The resulting electrode surfaces were characterized with both AFM and cyclic voltammetry. Anodic stripping voltammetry of arsenic(III) on the modified electrode was performed. After optimization, a LOD of 0.0096 ppb was obtained with LSV.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Oct
|
pubmed:issn |
0003-2700
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:day |
1
|
pubmed:volume |
76
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
5924-9
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading | |
pubmed:year |
2004
|
pubmed:articleTitle |
Anodic stripping voltammetry of arsenic(III) using gold nanoparticle-modified electrodes.
|
pubmed:affiliation |
Physical and Theoretical Chemistry Laboratory, University of Oxford, South Parks Road, Oxford OX1 3QZ, United Kingdom.
|
pubmed:publicationType |
Journal Article,
Research Support, Non-U.S. Gov't
|