Hypoxic mixtures of oxygen and nitrous oxide were demonstrated with three anaesthesia machines which had cyclopropane flowmeters. In one machine the bobbin did not prevent back flow and the hypoxic mixture occurred when the cyclopropane flow control was left open. Two other machines only delivered hypoxic mixtures if the cyclopropane bobbins were removed from their seats and the flow controls opened. The vaporizer on one machine increased the hypoxia when turned on.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0310-057X
pubmed:author
pubmed:issnType
Print
pubmed:volume
6
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
355-7
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
1978
pubmed:articleTitle
Hypoxia with a third flowmeter tube on the anaesthetic machine.