pubmed:abstractText |
A recoil-proton spectrometer was created by coupling a p-i-n diode with a polyethylene converter. The maximum detectable energy, imposed by the thickness of the totally depleted layer, is approximately 6 MeV. The minimum detectable energy is limited by the contribution of secondary electrons generated by photons in the detector assembly. This limit is approximately 1.5 MeV at full-depletion voltage and was decreased using pulse-shape discrimination. The diode was set up in the 'reverse-injection' configuration (i.e. with the N+ layer adjacent to the converter). This configuration provides longer collection times for the electron-hole pairs generated by the recoil-protons. The pulse-shape discrimination was based on the zero-crossing time of bipolar signals from a (CR)2-(RC)2 filter. The detector was characterised using monoenergetic neutrons generated in the Van De Graaff CN accelerator at the INFN-Laboratori Nazionali di Legnaro. The energy limit for discrimination proved to be approximately 900 keV.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:issn |
0144-8420
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
110
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
509-16
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:15353700-Body Burden,
pubmed-meshheading:15353700-Equipment Design,
pubmed-meshheading:15353700-Equipment Failure Analysis,
pubmed-meshheading:15353700-Feasibility Studies,
pubmed-meshheading:15353700-Humans,
pubmed-meshheading:15353700-Linear Energy Transfer,
pubmed-meshheading:15353700-Neutrons,
pubmed-meshheading:15353700-Protons,
pubmed-meshheading:15353700-Radiation Dosage,
pubmed-meshheading:15353700-Radiation Protection,
pubmed-meshheading:15353700-Radiometry,
pubmed-meshheading:15353700-Relative Biological Effectiveness,
pubmed-meshheading:15353700-Reproducibility of Results,
pubmed-meshheading:15353700-Risk Assessment,
pubmed-meshheading:15353700-Risk Factors,
pubmed-meshheading:15353700-Semiconductors,
pubmed-meshheading:15353700-Sensitivity and Specificity,
pubmed-meshheading:15353700-Signal Processing, Computer-Assisted,
pubmed-meshheading:15353700-Spectrum Analysis
|
pubmed:year |
2004
|
pubmed:articleTitle |
A recoil-proton spectrometer based on a p-i-n diode implementing pulse-shape discrimination.
|
pubmed:affiliation |
Dipartimento di Ingegneria Nucleare, Politecnico di Milano, via Ponzio 34/3, 20133 Milano, Italy. stefano.agosteo@polimi.it
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Research Support, Non-U.S. Gov't,
Evaluation Studies,
Validation Studies
|