pubmed:abstractText
Images of hyperpolarized 3He were acquired during breath-hold in four healthy volunteers with the use of an optimized 3D gradient-echo sequence. The images were compared with existing 2D gradient-echo methods. The average SNR from a 13-mm-thick slice in the peripheral lung was 1.4 times greater with 3D. In the airways the average SNR was 1.7 times greater with 3D. The higher SNR of 3D was particularly evident when regions of unimpeded gas diffusion, such as the major airways, were imaged with thin slices. This is because diffusion dephasing due to the slice-encoding gradient is minimized with a 3D sequence. The in vivo experimental findings were substantiated with experiments on phantoms of free gas, which showed more than four times the SNR with 3D compared to 2D. Theoretical simulations of the 2D and 3D k-space filters were also performed to predict the SNR and spatial resolution observed in the experimental images.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0740-3194
pubmed:author
pubmed:copyrightInfo
Copyright 2004 Wiley-Liss, Inc.
pubmed:issnType
Print
pubmed:volume
52
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
673-8
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2004
pubmed:articleTitle
Comparison between 2D and 3D gradient-echo sequences for MRI of human lung ventilation with hyperpolarized 3He.
pubmed:affiliation
Section of Academic Radiology, University of Sheffield, UK. j.m.wild@sheffield.ac.uk
pubmed:publicationType
Journal Article, Comparative Study