Septal penetration of high-energy photons may degrade the quality of single photon emission computed tomography (SPECT) of the heart with iodine 123-labeled tracers. We investigated the impact of collimator choice on cardiac SPECT with I-123.
|
pubmed:language |
eng
|
pubmed:journal |
|
pubmed:citationSubset |
IM
|
pubmed:chemical |
|
pubmed:status |
MEDLINE
|
pubmed:issn |
1071-3581
|
pubmed:author |
|
pubmed:issnType |
Print
|
pubmed:volume |
11
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
433-9
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:15295412-Cardiomyopathies,
pubmed-meshheading:15295412-Decision Support Systems, Clinical,
pubmed-meshheading:15295412-Equipment Failure Analysis,
pubmed-meshheading:15295412-Heart,
pubmed-meshheading:15295412-Humans,
pubmed-meshheading:15295412-Image Enhancement,
pubmed-meshheading:15295412-Iodine Radioisotopes,
pubmed-meshheading:15295412-Phantoms, Imaging,
pubmed-meshheading:15295412-Radiopharmaceuticals,
pubmed-meshheading:15295412-Reproducibility of Results,
pubmed-meshheading:15295412-Sensitivity and Specificity,
pubmed-meshheading:15295412-Technology Assessment, Biomedical,
pubmed-meshheading:15295412-Tomography, Emission-Computed, Single-Photon
|
pubmed:articleTitle |
Collimator choice in cardiac SPECT with I-123-labeled tracers.
|
pubmed:affiliation |
Department of Radiology, Institute of Medical Science, University of Tokyo, Japan. inoueys-tky@umin.ac.jp
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Evaluation Studies,
Validation Studies
|