Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
6
pubmed:dateCreated
2004-7-19
pubmed:abstractText
Design optimization, manufacturing, and tests, both laboratory and clinical, of a portable gamma camera for medical applications are presented. This camera, based on a continuous scintillation crystal and a position-sensitive photomultiplier tube, has an intrinsic spatial resolution of approximately 2 mm, an energy resolution of 13% at 140 keV, and linearities of 0.28 mm (absolute) and 0.15 mm (differential), with a useful field of view of 4.6 cm diameter. Our camera can image small organs with high efficiency and so it can address the demand for devices of specific clinical applications like thyroid and sentinel node scintigraphy as well as scintimammography and radio-guided surgery. The main advantages of the gamma camera with respect to those previously reported in the literature are high portability, low cost, and weight (2 kg), with no significant loss of sensitivity and spatial resolution. All the electronic components are packed inside the minigamma camera, and no external electronic devices are required. The camera is only connected through the universal serial bus port to a portable personal computer (PC), where a specific software allows to control both the camera parameters and the measuring process, by displaying on the PC the acquired image on "real time." In this article, we present the camera and describe the procedures that have led us to choose its configuration. Laboratory and clinical tests are presented together with diagnostic capabilities of the gamma camera.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0094-2405
pubmed:author
pubmed:issnType
Print
pubmed:volume
31
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1384-97
pubmed:dateRevised
2008-11-21
pubmed:meshHeading
pubmed:year
2004
pubmed:articleTitle
Design and tests of a portable mini gamma camera.
pubmed:affiliation
Instituto de Física Corpuscular (CSIC-UV), Edificio Institutos de Paterna, P.O. Box 22085, E-46071 Valencia, Spain. filomeno@ific.uv.es
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't, Evaluation Studies