We describe a case of potentially fatal undersensing of VF by a third generation ICD with predetermined automatic gain control. In this patient, ventricular sensing was optimal, as R wave amplitudes during sinus rhythm were at least 16 mV. Cyclical, high amplitude signals during VF elevated the sensing floor to such an extent that complete undersensing of subsequent lower amplitude local electrograms occurred. This led to bradypacing and complete ICD therapy failure. Therefore, high R wave amplitudes during sinus rhythm do not warrant flawless sensing during VF.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0147-8389
pubmed:author
pubmed:issnType
Print
pubmed:volume
27
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
833-4
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
2004
pubmed:articleTitle
Undersensing of VF in a patient with optimal R wave sensing during sinus rhythm.
pubmed:affiliation
Department of Cardiology, Academic Medical Center, Amsterdam, The Netherlands. L.R.Dekker@amc.uva.nl
pubmed:publicationType
Journal Article, Case Reports