The EyeSys Corneal Analysis System Model I measures corneal topography using digital image analysis of placido rings reflected off the cornea. With three observers, we compared the accuracy and reproducibility (precision) of the Marco Keratometer Model 1 and EyeSys Corneal Analysis System Model I using four poly(methyl methacrylate) spheres (37.50, 42.51, 47.54, and 55.06 diopters [D]), three steel spheres (40.50, 42.50, and 44.75 D), and 20 normal human eyes (41.50 to 46.00 D). For the spheres, the standard deviations of intra-observer and overall reproducibility for both devices were less than 0.12 D; the absolute mean differences between the measurements of the seven spheres and the known values were 0.25 D or more for two spheres as measured by the keratometer and none as measured by the EyeSys. For the normal corneas, the standard deviations of intra-observer and overall reproducibility for dioptric measurements were 0.07 D and 0.14 D for the keratometer and 0.13 D and 0.19 D for the EyeSys. The EyeSys Corneal Analysis System Model I exceeds keratometer accuracy in reading calibrated spheres and approaches keratometer reproducibility in measuring the 3-mm zone of normal human corneas.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Jul
|
pubmed:issn |
0886-3350
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
18
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
342-7
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:1501084-Cornea,
pubmed-meshheading:1501084-Humans,
pubmed-meshheading:1501084-Image Processing, Computer-Assisted,
pubmed-meshheading:1501084-Methylmethacrylates,
pubmed-meshheading:1501084-Observer Variation,
pubmed-meshheading:1501084-Ophthalmology,
pubmed-meshheading:1501084-Reproducibility of Results,
pubmed-meshheading:1501084-Sensitivity and Specificity,
pubmed-meshheading:1501084-Steel
|
pubmed:year |
1992
|
pubmed:articleTitle |
Comparison of the accuracy and reproducibility of the keratometer and the EyeSys Corneal Analysis System Model I.
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Research Support, Non-U.S. Gov't
|