Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
2-4
pubmed:dateCreated
2004-2-27
pubmed:abstractText
Since radionuclide calibrators are used in nuclear medicine for the determination of the activity of radiopharmaceuticals administered to patients, it is important to ensure their long-term accuracy. Results of the checks performed in the Czech Republic during the year 2002 are given, together with a short overview of previous years. For the CMI 4pigamma ionization chamber, against which all the calibrators are checked, we calculated the response function and basic characteristics by a Monte Carlo method, using the MCNP4C code.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
0969-8043
pubmed:author
pubmed:issnType
Print
pubmed:volume
60
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
535-8
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:articleTitle
Activity measurements with radionuclide calibrators in the Czech Republic.
pubmed:affiliation
Czech Metrology Institute, Inspectorate for Ionizing Radiation, Radiova 1, Prague, 102 00 Czech Republic. volsovcova@cmi.cz
pubmed:publicationType
Journal Article, Comparative Study, Evaluation Studies, Validation Studies