Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
5
pubmed:dateCreated
1993-2-17
pubmed:abstractText
Four commercial assay kits to measure serum ferritin were compared to establish the degree of agreement and interchangeability between the different techniques based on the intraclass correlation coefficient (ri). Radioimmunoassay, microparticle enzyme immunoassay, enzyme-linked immunosorbent assay, and chemiluminescent immunoassay systems were used. The Pearson product-moment correlation factor (r) between any two methods was at least 0.98, and the intercept of the regression equations ranged from -0.613 to 3.797, indicating that the methods were comparable. Furthermore, the intraclass correlation coefficient (ri) was at least 0.98, suggesting that the methods were interchangeable.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0002-9173
pubmed:author
pubmed:issnType
Print
pubmed:volume
98
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
511-5
pubmed:dateRevised
2011-11-17
pubmed:meshHeading
pubmed:year
1992
pubmed:articleTitle
Serum ferritin measurement and the degree of agreement using four techniques.
pubmed:affiliation
Department of Haematology, Adelaide Childrens Hospital, Australia.
pubmed:publicationType
Journal Article, Comparative Study