pubmed:abstractText |
Hydrogels containing benzo-18-crown-6 were used to modify microcantilevers for measurements of the concentration of Pb2+ in aqueous solutions. These microcantilevers undergo bending deflection upon exposure to solutions containing various Pb2+ concentrations as the result of a swelling of the hydrogels. It was found that a concentration of 10(-6) M Pb2+ can be detected using this technology. Other cations, such as Na+, have no effect on the deflection of this cantilever. The cation K+, which also complexes with benzo-18-crown-6, could interfere with Pb2+ detection, but only at high concentrations (> 10(-4) M).
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:status |
PubMed-not-MEDLINE
|
pubmed:month |
Jan
|
pubmed:issn |
0910-6340
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
20
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
9-11
|
pubmed:year |
2004
|
pubmed:articleTitle |
Detection of Pb2+ using a hydrogel swelling microcantilever sensor.
|
pubmed:affiliation |
Department of Chemistry and Institute for Micromanufacturing, Louisiana Tech University, Ruston, Louisiana 71272, USA.
|
pubmed:publicationType |
Journal Article
|