pubmed:abstractText |
Quantitative accuracy in iodine 123 studies may be impaired by septal penetration. We evaluated the effect of collimator choice on estimation of the heart-to-mediastinum (H/M) ratio in cardiac I-123 metaiodobenzylguanidine (MIBG) imaging.
|
pubmed:language |
eng
|
pubmed:journal |
|
pubmed:citationSubset |
IM
|
pubmed:chemical |
|
pubmed:status |
MEDLINE
|
pubmed:issn |
1071-3581
|
pubmed:author |
|
pubmed:issnType |
Print
|
pubmed:volume |
10
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
623-32
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:14668774-3-Iodobenzylguanidine,
pubmed-meshheading:14668774-Equipment Failure Analysis,
pubmed-meshheading:14668774-Extravasation of Diagnostic and Therapeutic Materials,
pubmed-meshheading:14668774-Heart Ventricles,
pubmed-meshheading:14668774-Image Enhancement,
pubmed-meshheading:14668774-Mediastinum,
pubmed-meshheading:14668774-Phantoms, Imaging,
pubmed-meshheading:14668774-Radiometry,
pubmed-meshheading:14668774-Radiopharmaceuticals,
pubmed-meshheading:14668774-Reproducibility of Results,
pubmed-meshheading:14668774-Scattering, Radiation,
pubmed-meshheading:14668774-Sensitivity and Specificity,
pubmed-meshheading:14668774-Sympathetic Nervous System,
pubmed-meshheading:14668774-Tomography, Emission-Computed
|
pubmed:articleTitle |
Effect of collimator choice on quantitative assessment of cardiac iodine 123 MIBG uptake.
|
pubmed:affiliation |
Department of Radiology, Institute of Medical Sciences, University of Tokyo, Japan. inoueys-tky@umin.ac.jp
|
pubmed:publicationType |
Journal Article,
Comparative Study,
Evaluation Studies,
Validation Studies
|