pubmed:abstractText |
We describe a rapid and sensitive latex nephelometric immunoassay for quantifying ferritin in human serum. This latex immunoassay procedure uses commercially available ready-for-use reagents [Tina-Quant (a) Ferritin, Boehringer Mannheim] that have a long shelf life. The assay consists of incubating the diluted serum sample (5-fold) for 12 min at room temperature with latex particles covalently coated with anti-ferritin antibodies, and then quantifying the change of light-scatter produced. The assay is fully automated on the Behring nephelometer analyzer with a sampling rate of 150 samples/hour. The method has an analytical range of 3 to 260 micrograms/l. Maximal intra- and inter-assay CVs were 4.0 and 6.2%, respectively. Analytical recoveries ranged from 91.3 to 103.6%. Assay detection limit was less than 3 micrograms/l. Linearity of the test is given throughout the measuring range. There was no interference from bilirubin (up to 340 mumol/l), haemoglobin (up to 7 g/l), or rheumatoid factor (up to 1,100 IU/ml). Turbid and lipemic samples interfere. This interference may be avoided by pretreating these samples prior to assay. Results correlated well with those obtained by an automated ELISA test (r = 0.995) and with those of two commercial RIA methods (r greater than 0.97). This latex nephelometric procedure is a convenient method and represents an interesting alternative to other immunoassays for measuring ferritin in human serum.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:issn |
0887-8013
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
6
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
239-44
|
pubmed:dateRevised |
2007-11-15
|
pubmed:meshHeading |
pubmed-meshheading:1403343-Adolescent,
pubmed-meshheading:1403343-Adult,
pubmed-meshheading:1403343-Evaluation Studies as Topic,
pubmed-meshheading:1403343-Female,
pubmed-meshheading:1403343-Ferritins,
pubmed-meshheading:1403343-Humans,
pubmed-meshheading:1403343-Latex Fixation Tests,
pubmed-meshheading:1403343-Male,
pubmed-meshheading:1403343-Middle Aged,
pubmed-meshheading:1403343-Nephelometry and Turbidimetry,
pubmed-meshheading:1403343-Reference Values,
pubmed-meshheading:1403343-Sensitivity and Specificity
|
pubmed:year |
1992
|
pubmed:articleTitle |
Automated quantitative nephelometric latex immunoassay for determining ferritin in human serum.
|
pubmed:affiliation |
Laboratorio Central, Hospital San Millan, Logroño, Spain.
|
pubmed:publicationType |
Journal Article,
Comparative Study
|