pubmed:abstractText |
An HPGe gamma probe designed to monitor radioactive nuclides in environmental water in situ is described. The probe is equipped with a 15% HPGe detector and an associated spectrum analyzer. Laboratory tests were performed to evaluate its operating depth, detecting sensitivity, detecting volume and the detection limits of radionuclides. A field operation was conducted to measure in situ radionuclide concentrations in a nuclear reactor pool, and the feasibility as well as the disadvantages of this rapid survey are discussed.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Apr
|
pubmed:issn |
0883-2889
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
43
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
475-80
|
pubmed:dateRevised |
2009-6-4
|
pubmed:meshHeading | |
pubmed:year |
1992
|
pubmed:articleTitle |
Radionuclide monitoring in environmental water body using an in situ gamma probe.
|
pubmed:affiliation |
Nuclear Science and Technology Development Center, National Tsing Hua University, Hsinchu, Taiwan, Republic of China.
|
pubmed:publicationType |
Journal Article,
Research Support, Non-U.S. Gov't
|