pubmed:abstractText
A method for estimating the characteristics of an automatic matching control (AMC) system for in vivo electron paramagnetic resonance (EPR) spectroscopy is described. AMC is also known as automatic coupling control (ACC) and is used to compensate for perturbation due to movement of the subject animal. The open-loop function of the AMC system is derived and the characteristics (gain, phase margin, steady-state error, system bandwidth, and disturbance rejection) of the system are examined. The proposed method is supported by experimental findings. The frequency response of perturbation due to the movement of an anesthetized rat was measured with a 1.1-GHz continuous-wave EPR spectrometer. Disturbance rejection with the AMC system was also considered for respiratory motion and the heartbeat of the anesthetized rat, and these perturbations were reduced by about 40 dB in the power spectrum of the output of the diode detector in the spectrometer.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jul
pubmed:issn
0740-3194
pubmed:author
pubmed:copyrightInfo
Copyright 2003 Wiley-Liss, Inc.
pubmed:issnType
Print
pubmed:volume
50
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
223-7
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed-meshheading:12815701-Abdomen, pubmed-meshheading:12815701-Animals, pubmed-meshheading:12815701-Artifacts, pubmed-meshheading:12815701-Electron Spin Resonance Spectroscopy, pubmed-meshheading:12815701-Electronics, pubmed-meshheading:12815701-Equipment Design, pubmed-meshheading:12815701-Equipment Failure Analysis, pubmed-meshheading:12815701-Feedback, pubmed-meshheading:12815701-Female, pubmed-meshheading:12815701-Image Enhancement, pubmed-meshheading:12815701-Movement, pubmed-meshheading:12815701-Quality Control, pubmed-meshheading:12815701-Rats, pubmed-meshheading:12815701-Rats, Wistar, pubmed-meshheading:12815701-Reproducibility of Results, pubmed-meshheading:12815701-Respiratory Mechanics, pubmed-meshheading:12815701-Sensitivity and Specificity, pubmed-meshheading:12815701-Thorax
pubmed:year
2003
pubmed:articleTitle
Control characteristics of an automatic matching control system for in vivo EPR spectroscopy.
pubmed:affiliation
Department of Electrical Engineering, Yamagata University, Yonezawa, Japan. hhirata@yz.yamagata-u.ac.jp
pubmed:publicationType
Journal Article, Comparative Study, Research Support, Non-U.S. Gov't, Evaluation Studies, Validation Studies