A method to establish the relationship between CT number and effective density for therapeutic radiations is proposed. We approximated body tissues to mixtures of muscle, air, fat and bone. Consequently, the relationship can be calibrated only with a CT scan of their substitutes, for which we chose water, air, ethanol and potassium phosphate solution, respectively. With simple and specific corrections for non-equivalencies of the substitutes, a calibration accuracy of 1% will be achieved. We tested the calibration method with some biological materials to verify that the proposed method would offer the accuracy, simplicity and specificity required for a standard in radiotherapy treatment planning, in particular with heavy charged particles.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0031-9155
pubmed:author
pubmed:issnType
Print
pubmed:day
21
pubmed:volume
48
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1053-64
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2003
pubmed:articleTitle
A CT calibration method based on the polybinary tissue model for radiotherapy treatment planning.
pubmed:affiliation
Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, 4-9-1 Anagawa, Inage-Ku, Chiba 263-8555, Japan. nkanemat@nirs.go.jp
pubmed:publicationType
Journal Article, Comparative Study, Evaluation Studies, Validation Studies