pubmed:abstractText |
A 71-year-old woman had undergone valvular heart surgery in 1981, and implantation of a permanent ventricular pacemaker for ventricular pauses during atrial fibrillation in 2001. One year after pacemaker implantation, she complained of faintness. When pacing at 100 beats/min the pacemaker functioned properly. However, pacing and sensing failure was detected at a pacing rate of 60 beats/min. At rapid pacing rates, the lead tip was in closer contact with the endocardium, and its microdislodgment was undetectable. The symptoms have resolved since the lead was repositioned.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Mar
|
pubmed:issn |
0147-8389
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
26
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
787-8
|
pubmed:dateRevised |
2004-11-17
|
pubmed:meshHeading |
pubmed-meshheading:12698687-Aged,
pubmed-meshheading:12698687-Atrial Fibrillation,
pubmed-meshheading:12698687-Cardiac Pacing, Artificial,
pubmed-meshheading:12698687-Electrocardiography,
pubmed-meshheading:12698687-Electrodes, Implanted,
pubmed-meshheading:12698687-Equipment Failure,
pubmed-meshheading:12698687-Female,
pubmed-meshheading:12698687-Heart Rate,
pubmed-meshheading:12698687-Humans,
pubmed-meshheading:12698687-Pacemaker, Artificial,
pubmed-meshheading:12698687-Time Factors
|
pubmed:year |
2003
|
pubmed:articleTitle |
Microdislodgment of ventricular pacing lead undetectable during rapid pacing one year after implantation.
|
pubmed:affiliation |
First Department of Internal Medicine, Niigata University School of Medicine, 2-746 Asahimachi, Niigata 951-8518, Japan.
|
pubmed:publicationType |
Journal Article,
Case Reports
|