Information on electron energy is important in planning radiation therapy using electrons. The Geske 3405 electron beam energy monitor (Geske monitor, PTW Nuclear Associates, Carle Place, NY, USA) is a device containing nine ionization chambers for checking the energy of the electron beams produced by radiotherapy accelerators. We wondered whether this might increase the likelihood of ionization chamber trouble. In spite of the importance of the stability of such a quality assurance (QA) device, there are no reports on the stability of values measured with a Geske monitor. The purpose of this paper was therefore to describe the stability of a Geske monitor. It was found that the largest coefficient of variation (CV) of the Geske monitor measurements was approximately 0.96% over a 21-week period. In conclusion, the stability of Geske monitor measurements of the energy of electron beams from a linear accelerator was excellent.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0040-8727
pubmed:author
pubmed:issnType
Print
pubmed:volume
198
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
197-201
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
2002
pubmed:articleTitle
Stability of electron-beam energy monitor for quality assurance of the electron-beam energy from radiotherapy accelerators.
pubmed:affiliation
Department of Radiological Technology, College of Medical Sciences, Tohoku University, Sendai 980-8575, Japan. chida@rad.cms.tohoku.ac.jp
pubmed:publicationType
Journal Article