The good biomechanical and tribological properties, together with the excellent biocompatibility, of ceramic-on-ceramic components, make them a preferential choice for total hip replacement surgery, at least in Europe. We report on a man admitted as an outpatient with painless grating in the hip one year after replacement of a ceramic femoral head, but not of the ceramic inlay. Clinical and radiological findings were indicative of a broken liner. This was confirmed during revision surgery, during which it was replaced by a polyethylene inlay; although the ceramic head appeared intact, it was replaced by a metal head. Inspection of the surface of the broken liner in the scanning electron microscope (SEM) revealed signs of material failure. We recommend careful inspection of ceramic-on-ceramic articulating components during total hip revision surgery and if there is any uncertainty, replacement of both so as to avoid premature failure.
pubmed:language
ger
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
0013-5585
pubmed:author
pubmed:issnType
Print
pubmed:volume
47
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
306-9
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2002
pubmed:articleTitle
[Fracture analysis of a ceramic liner. Is in hip endoprosthesis replacement of ceramic on ceramic components with only one of the corresponding partners justified?].
pubmed:affiliation
Orthopädische Universitätsklinik Düsseldorf, Moorenstrasse 5, D-40225 Düsseldorf. Jaeger@med.uni-duesseldorf.de
pubmed:publicationType
Journal Article, Comparative Study, English Abstract, Case Reports