pubmed:abstractText |
The good biomechanical and tribological properties, together with the excellent biocompatibility, of ceramic-on-ceramic components, make them a preferential choice for total hip replacement surgery, at least in Europe. We report on a man admitted as an outpatient with painless grating in the hip one year after replacement of a ceramic femoral head, but not of the ceramic inlay. Clinical and radiological findings were indicative of a broken liner. This was confirmed during revision surgery, during which it was replaced by a polyethylene inlay; although the ceramic head appeared intact, it was replaced by a metal head. Inspection of the surface of the broken liner in the scanning electron microscope (SEM) revealed signs of material failure. We recommend careful inspection of ceramic-on-ceramic articulating components during total hip revision surgery and if there is any uncertainty, replacement of both so as to avoid premature failure.
|
pubmed:language |
ger
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Dec
|
pubmed:issn |
0013-5585
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
47
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
306-9
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:12585047-Aged,
pubmed-meshheading:12585047-Aluminum Oxide,
pubmed-meshheading:12585047-Equipment Failure Analysis,
pubmed-meshheading:12585047-Hip Prosthesis,
pubmed-meshheading:12585047-Humans,
pubmed-meshheading:12585047-Male,
pubmed-meshheading:12585047-Materials Testing,
pubmed-meshheading:12585047-Microscopy, Electron, Scanning,
pubmed-meshheading:12585047-Ossification, Heterotopic,
pubmed-meshheading:12585047-Postoperative Complications,
pubmed-meshheading:12585047-Prosthesis Design,
pubmed-meshheading:12585047-Prosthesis Failure,
pubmed-meshheading:12585047-Reoperation,
pubmed-meshheading:12585047-Surface Properties
|
pubmed:year |
2002
|
pubmed:articleTitle |
[Fracture analysis of a ceramic liner. Is in hip endoprosthesis replacement of ceramic on ceramic components with only one of the corresponding partners justified?].
|
pubmed:affiliation |
Orthopädische Universitätsklinik Düsseldorf, Moorenstrasse 5, D-40225 Düsseldorf. Jaeger@med.uni-duesseldorf.de
|
pubmed:publicationType |
Journal Article,
Comparative Study,
English Abstract,
Case Reports
|