pubmed:abstractText |
A time-resolved acquisition technique has been investigated for spatially-resolved electron energy-loss spectroscopy by energy-fitering transmission electron microscopy. Plural (10) spectral images were acquired separately at an energy (E) and a position (Y) coordinate. After correction of both specimen and energy drifts by comparing with the original starting spectrum, the 10 spectral images were added to form a new processed spectral image. This time-resolved acquisition technique was found to improve the signal-to-noise ratio to about the square root of ten for 200 s total exposure time without increasing the image blurring problem.
|
pubmed:language |
eng
|
pubmed:status |
PubMed-not-MEDLINE
|
pubmed:author | |
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
291-6
|
pubmed:dateRevised |
2003-11-4
|
pubmed:articleTitle |
Time-resolved acquisition technique for spatially-resolved electron energy-loss spectroscopy by energy-filtering TEM.
|