Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
3
pubmed:dateCreated
2002-4-30
pubmed:abstractText
Miniature analyzer Glucophot BP-"M", intended for express measurements of glucose concentrations in capillary blood, was tested under clinical conditions. The reference method was the hexokinase method for glucose measurements on a Spectrum 11 biochemical analyzer (Abbott); in addition, glucose measurements were carried out on a One Touch-11 glucometer (Johnson-Johnson). The results evidence that the reproducibility and correctness of glucose analysis by the Glucophot BP-"M" analyzer both in health and disease met the requirements to laboratory measurements performed by "dry" chemical methods and demonstrated its high analytical reliability.
pubmed:language
rus
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Mar
pubmed:issn
0869-2084
pubmed:author
pubmed:issnType
Print
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
9-10
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2002
pubmed:articleTitle
[Glucophot BP-"M" analyzer for blood analysis].
pubmed:publicationType
Journal Article, English Abstract