Source:http://linkedlifedata.com/resource/pubmed/id/11934173
Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
4
|
pubmed:dateCreated |
2002-4-5
|
pubmed:abstractText |
X-ray imaging in differential interference contrast (DIC) with submicrometer optical resolution was performed by using a twin zone plate (TZP) setup generating focal spots closely spaced within the TZP spatial resolution of 160 nm. Optical path differences introduced by the sample are recorded by a CCD camera in a standard full-field imaging and by an aperture photodiode in a standard scanning transmission x-ray microscope. Applying this x-ray DIC technique, we demonstrate for both the full-field imaging and scanning x-ray microscope methods a drastic increase in image contrast (approximately 20x) for a low-absorbing specimen, similar to the Nomarski DIC method for visible-light microscopy.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:status |
PubMed-not-MEDLINE
|
pubmed:month |
Apr
|
pubmed:issn |
1084-7529
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
19
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
797-806
|
pubmed:dateRevised |
2003-11-3
|
pubmed:year |
2002
|
pubmed:articleTitle |
Differential interference contrast x-ray microscopy with twin zone plates.
|
pubmed:affiliation |
X-Ray Microscopy Section, ELETTRA-Sincrotrone Trieste, Basovizzo-Trieste, Italy. kaulich@elettra.trieste.it
|
pubmed:publicationType |
Journal Article
|