pubmed:abstractText
To study gas exchange in the filter during continuous venovenous hemodiafiltration (CVVHDF), an air-tight heated mixing chamber with adjustable CO2 supply was constructed and connected to a CVVHDF monitor. Bicarbonate-free crystalloid (Part 1) and packed red blood cell (Part 2) solutions were circulated at 150 ml x min(-1). Gas exchange expressed as pre-postfilter difference in CO2 and O2 contents was measured at different CVVHDF settings and temperatures of circulating and dialysis solutions. Ultrafiltration was most efficacious for CO2 removal (at 1,000 ml x h(-1) ultrafiltration CO2 losses reached 13% of prefilter CO2 content). Addition of dialysis (1,000 ml x h(-1)) increased CO2 loss to 17% and at maximal parameters (filtration 3,000 ml x h(-1), dialysis 2,500 ml x h(-1)), the loss of CO2 amounted to 35% of prefilter content. Temperature changes of circulating and/or dialysis fluids had no significant impact on CO2 losses. The O2 exchange during CVVHDF was negligible. Currently used CVVHDF is only marginally effective in CO2 removal. Higher volume ultrafiltration combined with dialysis can be expected to reach clinical significance.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
0160-564X
pubmed:author
pubmed:issnType
Print
pubmed:volume
25
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
961-6
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2001
pubmed:articleTitle
Effects of ultrafiltration, dialysis, and temperature on gas exchange during hemodiafiltration: a laboratory experiment.
pubmed:affiliation
Department of Biophysics, Medical Faculty of Charles University, Pilsen, Czech Republic. ruzicka@dante.lfp.cuni.cz
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't