pubmed:abstractText |
A case of near fatal obstruction of a breathing system is described, due to occlusion of a heat and moisture exchange filter in a patient with pulmonary edema. Previous reports have described cases of mechanical obstruction to fresh gas flow due to an accumulation of a bolus of fluid within the filter housing. However, in this case, the cause of occlusion was due to the protein and cell debris contained in a small amount of pulmonary edema fluid interacting with the filter membrane.
|
pubmed:commentsCorrections |
|
pubmed:language |
eng
|
pubmed:journal |
|
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Feb
|
pubmed:issn |
0832-610X
|
pubmed:author |
|
pubmed:issnType |
Print
|
pubmed:volume |
49
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
126-31
|
pubmed:dateRevised |
2004-11-17
|
pubmed:meshHeading |
|
pubmed:year |
2002
|
pubmed:articleTitle |
Rapid and complete occlusion of a heat and moisture exchange filter by pulmonary edema (clinical report).
|
pubmed:affiliation |
Department of Anaesthesia and Intensive Care, University Hospital of Wales, Heath Park, Cardiff, United Kingdom. davidjwilliams@doctors.org.uk
|
pubmed:publicationType |
Journal Article,
Case Reports
|