pubmed:abstractText
A case of near fatal obstruction of a breathing system is described, due to occlusion of a heat and moisture exchange filter in a patient with pulmonary edema. Previous reports have described cases of mechanical obstruction to fresh gas flow due to an accumulation of a bolus of fluid within the filter housing. However, in this case, the cause of occlusion was due to the protein and cell debris contained in a small amount of pulmonary edema fluid interacting with the filter membrane.
pubmed:commentsCorrections
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Feb
pubmed:issn
0832-610X
pubmed:author
pubmed:issnType
Print
pubmed:volume
49
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
126-31
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
2002
pubmed:articleTitle
Rapid and complete occlusion of a heat and moisture exchange filter by pulmonary edema (clinical report).
pubmed:affiliation
Department of Anaesthesia and Intensive Care, University Hospital of Wales, Heath Park, Cardiff, United Kingdom. davidjwilliams@doctors.org.uk
pubmed:publicationType
Journal Article, Case Reports