rdf:type |
|
lifeskim:mentions |
|
pubmed:issue |
12
|
pubmed:dateCreated |
2002-1-30
|
pubmed:abstractText |
A 63-year-old man with chronic atrial fibrillation and heart failure had a biventricular pacing system implanted. The pulse generator was a standard DDDR pacemaker, using the atrial channel for the right ventricular lead and the ventricular channel for the left ventricular lead. During final adjustment of the pacing parameters, a pacemaker tachycardia triggered by T wave oversensing from the right ventricular lead was recorded.
|
pubmed:commentsCorrections |
|
pubmed:language |
eng
|
pubmed:journal |
|
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Dec
|
pubmed:issn |
0147-8389
|
pubmed:author |
|
pubmed:issnType |
Print
|
pubmed:volume |
24
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1819-20
|
pubmed:dateRevised |
2007-11-15
|
pubmed:meshHeading |
pubmed-meshheading:11817819-Atrial Fibrillation,
pubmed-meshheading:11817819-Cardiac Pacing, Artificial,
pubmed-meshheading:11817819-Electrocardiography,
pubmed-meshheading:11817819-Equipment Design,
pubmed-meshheading:11817819-Heart Failure,
pubmed-meshheading:11817819-Humans,
pubmed-meshheading:11817819-Male,
pubmed-meshheading:11817819-Middle Aged,
pubmed-meshheading:11817819-Pacemaker, Artificial,
pubmed-meshheading:11817819-Tachycardia
|
pubmed:year |
2001
|
pubmed:articleTitle |
Pacemaker-mediated tachycardia in a biventricular pacing system.
|
pubmed:affiliation |
Department of Cardiology, Catharina Hospital, Eindhoven, The Netherlands.
|
pubmed:publicationType |
Journal Article,
Case Reports
|