pubmed:abstractText
Modern ventilators are complicated electronic instruments with microprocessors and software, with the possibility of technical errors and problems such as autocycling. Despite autocycling being recognized as a problem in textbooks and reviews, there are few reports about autocycling in the literature. We report a case where a sudden increase in respiratory frequency due to autocycling resulted in a dangerous increase in intrinsic positive end-expiratory pressure (intrinsic PEEP, PEEPi). We think our case illustrates that autocycling does occur, but that the exact underlying mechanism may be hard to document and understand for clinicians. To remedy this situation, we suggest that manufacture-independent technical expertise should be established to evaluate incidents and suggest improvements.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0001-5172
pubmed:author
pubmed:issnType
Print
pubmed:volume
45
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1295-7
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
2001
pubmed:articleTitle
Autocycling and increase in intrinsic positive end-expiratory pressure during mechanical ventilation.
pubmed:affiliation
Department of Anaesthesia and Intensive Care, Rogaland Central Hospital, Stavanger, Norway. svein@harboe.org
pubmed:publicationType
Journal Article, Case Reports