pubmed:abstractText
This paper describes simple methods for determining spectral bandwidth by measurement and by calculation. We measured the spectral bandwidth of the monochromator from the Beckman Enzyme Analyzer TR, and found it to be 3.5-4.0 nm (av 3.8) by two different methods. We visually estimated it to be 4 nm. Using the specifications for the diffraction grafting, focal length, and fixed slit width, we calculated the theoretical spectral bandwidth to be 3.6 nm.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
0009-9147
pubmed:author
pubmed:issnType
Print
pubmed:volume
21
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1582-4
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1975
pubmed:articleTitle
Measurement of spectral bandwidth, as exemplified with the Beckman "Enzyme Analyzer System TR spectrophotometer".